The stability chambers specifications required from most of our pharmaceutical customers stability chambers, are designed for 25°C, 60% relative humidity and /or 40°C, 75% relative humidity.
SCS stability chambers can provide accurate and constant temperature and humidity control and uniformity to meet your specifications
SCS stability chambers are designed for longterm, accelerated and photostability studies
SCS can design/build single compartment to multi compartment stability chambers
SCS provides full service validation of your stability chamber including the following:
Installation qualification (IQ): confirms that all the stability chambers were installed to specifications. All of the model and serial numbers are documented
Operational qualification (OQ): verification of operator controls, alarms, empty stability chamber temperature mapping and verification that chamber stays within the humidity tolerances. OQ testing also includes temperataure recovery test for door openings and temmperature decay test to determine how long the chamber will remain within specifications due to power failure
Performance qualification (PQ): confirms that the stability chambers are meeting the operating setpoints at simulated product load. The temperature and humidity uniformities are also confirmed
Sample Validation Protocols available for your review